Pressure types |
Gauge pressure |
Measuring range |
10kPa-3MPa, please refer to the
ordering information chapter |
Output signal |
4-20mA, 4-20mA+HART, customer |
Reference accuracy |
±0.2% URL |
|
Stability: ±0.2% URL/year |
Including linearity, hysteresis and repeatability.
calibration temperature: 20°C±5°C |
|
Linear
output
accuracy Typical |
±0.2% URL , Nominal value
40kPa、250kPa
1MPa、3MPa |
Linear
output
accuracy Max value |
±0.5% URL, Nominal value
40kPa、250kPa
1MPa、3MPa |
The accuracy of square root output is 1.5 times of above
linear reference output accuracy. |
The overall performance including but not limited to
【Reference accuracy】,【Environment temperature
effects】, 【Static pressure effects】and other
comprehensive error |
Output signal |
|
Signal |
4-20mA |
Type |
Linearity |
Output |
Output |
Signal |
4-20mA+HART |
Type |
Linearity |
Output |
Two wire |
Ambient temperature effects |
|
Within the range -20-80°C total impact |
±0.2% URL/10K |
Power supply effects |
Zero and span change should not be more than ± 0.005%
URL/V when power supply changes in 10.5/16.5-55VDC |
Loading effects |
Zero and span change should not be more than ± 0.05%
URL/kΩ |
Vibration effects |
According to IEC60068-2-6 , 10g RMS (25-
2000HZ) |
Impact
resistence |
According to IEC60068-2-27 , 500g/1ms |
Insulation resistance |
≥ 20M Ω@ reference, 100VDC |
Power supply |
|
Item |
Operating conditions |
Standard |
10.5-55VDC |
HART protocol |
16.5-55VDC, communication
load resistance 250Ω |
Load resistance |
0-2119Ω for operation,
250-600Ω for HART protocol |
Transmission distance |
<1000 meters |
Power consumption |
≤500mW@24VDC, 20.8mA |
Working
temperature |
-40-85°C, integrated LCD display:
-20-70°C |
Storage temperature |
-40-110°C, integrated LCD display:
-40-85°C |
Media temperature |
Hygienic fluid filling: -10-125°C;
with heat exchange connector:
-10-250°C* |
|
Silicon oil filling: -40-120°C,
with heat exchange connector:
-40-300°C* |
Working humidity |
0-95%RH |
Proction class |
IP67 |
Dangerous condition |
ExiaIICT4(GYB16.1965X)** |
Test conditions |
30MHz-1000MHz |
|
0.15MHz-30MHz |
Measuring medium |
|
viscous, paste-like, adhesive, crystallising,
particulatescontaining and contaminated media |
Weight |
Net weight: about 0.6kg ( without mounting bracket and
process connection adaptor ) |